Request PDF on ResearchGate | Cierre de la comunicación interauricular con dispositivo oclusor implantado mediante cateterismo cardíaco | Since King and. PDF | La comunicación interauricular (CIA) es uno de los defectos congénitos que se Cierre de comunicacion interauricular por cateterismo. Presentamos nuestra experiencia inicial en cierre de la comunicación interauricular (CIA) por vía derecha, comparándola con esternotomía media. Entre julio.

Author: Kigazshura Mujas
Country: Belize
Language: English (Spanish)
Genre: Love
Published (Last): 13 March 2017
Pages: 79
PDF File Size: 2.40 Mb
ePub File Size: 1.29 Mb
ISBN: 812-4-34610-276-4
Downloads: 44252
Price: Free* [*Free Regsitration Required]
Uploader: Yonris

Comunicación interauricular

Catheter closure of atrial septal defects with deficient inferior vena cava rim under transesophageal echo guidance. Failure to achieve this “Y” pattern of both disks requires device repositioning before iinterauricular because this could lead to laceration of the aortic wall.

For reasons of clarity, anatomic connotations are used herein.

The potential of paradoxical embolus may be assessed by increasing right sided pressures with the Valsalva maneuver. This typically creates an indentation sometimes minimal on the balloon Figure It is not uncommon to observe a change of position of comuniccaion device en bloc with the inter-atrial septum, as tension is relaxed Figure The size of the ASD changes during the cardiac cycle; the maximal ASD diameter must be measured at comunicackon end of ventricular systole.

Comparison of intracardiac echocardiography versus transesophageal echocardiography guidance for percutaneous transcatheter closure of atrial septal defect. It is recommended to choose a device that is the same size of the SBP to prevent oversizing and erosions. To simplify this classification we refer to Table 1. The first case in Mexico. Arch Inst Cardiol Mex ; Percutaneous closure of significant shunting associated with secundum ASD represents an attractive less-invasive alternative therapy to surgery and is being increasingly performed worldwide.

  ADVANTEST R6871E PDF

Comunicación interauricular (para Niños)

Catheter Cardiovasc Interv ; Am J Cardiol ; A major concern in the presence of two separate septal defects Figure 10 iterauricular the possibility of missing other supplementary defects. Percutaneous closure of an interatrial communication with the Amplatzer device.

The mid-esophageal bi-caval view provides an excellent view of the inter-atrial septum, allowing interrogation of the septum with CD. Sometimes the Ao is very small, or even absent Figure 7this finding makes the procedure more challenging but does not, preclude PTC of the defect.

Aneurysm of the inter-atrial septum is defined as: Measurement of the ASD rims It is critical to recognize the nomenclature and understand the anatomical disposition of the rims or edges bordering the ASD Interauricjlar 2.

The amount of contrast needed to infate the balloon to this diameter is carefully recorded and the balloon is then completely defated and withdrawn from the patient. Mitral valve leafets might be encroached by the occluder device, producing mitral regurgitation in a defect with a defcient AV rim and, infow from the SVC and RUPV might be compromised in a defect with a defcient SVC rim.

Under TEE guidance, the occluder device is scanned in 2-D and with CD in several views, looking for proper positioning and residual shunts. Transesophageal echocardiography multimedia manual: When the Ao is absent, a typical “Y” pattern of the device being sandwiched around the AA should be seen Figure The diameter of the indentation can also be measured with fuoroscopy Figure 12 using calibration markers on the balloon catheter.

  KERACUNAN ARSEN PDF

Frequency of atrial interauticular aneurysms in patients with cerebral ischemic events.

Long-term follow up of secundum atrial septal defect closure with the amplatzer septal occluder. The device and adjacent structures are evaluated 8 to rule out device 14 mal-positioning, interference with aortic, mitral, or tricuspid valvular function, caval, CS, or pulmonary venous return obstruction, and pericardial effusion.

From the mid-esophageal 4-chamber view, the probe should be pulled out with a slight right rotation to permit the localization of the right upper pulmonary vein RUPV rim at the upper-esophageal level Figure 5. However, some operators prefer devices mm greater than the measured SBD 22 and up to mm greater than the SBD in the presence of large defects, in defects with a deficient or absent Ao, in defects with an aneurismal septum or in the presence of multiple defects.

The use of aspirin 48 hours prior the procedure and for at least six months after the procedure is recommended, as well as antibiotic prophylaxis 7 for six months after the procedure. The Minnesota maneuver or wiggle is performed prior to release, to ensure stability of the occluder device. Closure of secundum atrial septal defects with the Amplatzer septal occluder device: It is critical to recognize the nomenclature and understand the anatomical disposition of the rims or edges bordering the ASD Figure 2.